12/11/2026
Load - Seal – Perfuse
The idea behind the DynamicOrgan® Load&Seal Kit is simple: load first, seal afterwards, then perfuse.
With the open-top biochip, researchers can place precision-cut tissue slices, biomaterials, non-flowable samples or larger constructs directly into the top cell culture chamber before closing the system with the bonding foil.
1. Step – Load
After sterilization of both the biochip and separate, top bonding foil, add a small drop of medium or DPBS to the membrane and then place your tissue slice, drug depot other large biological sample in the top channel.
2. Step – Seal
Remove the protective film from the biochip and bonding foil. Then hold the top foil with both hands and slowly position it in the center of the biochip, making sure to align the foil’s cutouts precisely with the biochip ports.
3. Step – Perfuse
After application of the bonding foil, both top and bottom channel can be perfused with individual flow rates or no flow at all such as in an air-liquid interface.